Quick Enquiry

Indicators

Indicators

All Services

Equipment Details :

  • Make : Beamex
  • Model No : MC 2

Technical Details

Used for simulation and measurement of temperature of various types of thermocouples and RTDs, DC Voltage ,DC Current, Resistance etc

Accuracy :

Measurement:

  • Voltage : 0.02% RDG+ 0.25 mV
  • Current : 0.02% RDG +1.5 µA
  • Frequency : 0.01% RDG
  • Resistance : 0.02% RDG +3.5 mΩ
  • Thermocouples: ±0.0015 %
  • RTD : ±0.0015 %

Simulation :

  • Voltage : 0.02% RDG +0.1mV
  • Current : 0.02% RDG +1.5 µA
  • Resistance : 0.04% RDG
  • Frequency : 0.01% RDG
  • Thermocouples: ±0.0015 %
  • RTD : ±0.0015 %

Application

Calibration of furnace using Thermcouple with Multifunction calibrator :

Calibration of Temperature controller /Indicator:

Equipment Details :

  • Make : Yokogawa
  • Model No : MV 1000

Technical Details

16 Channel Data Logger Used for continous measurement of Temperature,Voltage ,Humidity

Accuracy :

  • DC Voltage : 0.05%
  • RTD : 0.15% of rdg + 0.3° C
  • Thermocouple : 0.15% of rdg + 1°C (R,S,B Type) : 0.15% of rdg + 0.7°C (K Type) : 0.15% of rdg +0.5°C (E,J,T Type )

Application

Calibration of Oven/Chamber/Furnace using Thermcouples with Paperless recorder :